污水處理設備 污泥處理設備 水處理過濾器 軟化水設備/除鹽設備 純凈水設備 消毒設備|加藥設備 供水/儲水/集水/排水/輔助 水處理膜 過濾器濾芯 水處理濾料 水處理劑 水處理填料 其它水處理設備
廈門慧嘉生物科技有限公司
暫無信息 |
閱讀:256發布時間:2011-8-26
小鼠(INS)小鼠胰島素(INS)ELISA試劑盒說明書
----------------------- Page 1-----------------------
Mouse Insulin (INS)ELISA Kit
Catalog No. CSB-E05071m
(96T)
This immunoassay kit allows for the in vitro quantitative determination of mouse INS
concentrations in serum, plasma and other biological fluids.
Expiration date six months from the date of manufacture
FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
1
----------------------- Page 2-----------------------
INTRODUCTION
Insulin is a peptide hormone composed of 51 amino acid residues and has a
molecular weight of 5808 Da. It is produced in the islets of Langerhans in the
pancreas. Insulin's structure varies slightly between species of animal. Insulin
has extensive effects on both metabolism and several other body systems (eg,
vascular compliance). Insulin causes most of the body's cells to take up
glucose from the blood (including liver, muscle, and fat tissue cells), storing it
as glycogen in the liver and muscle, and stops use of fat as an energy source.
When insulin is absent (or low), glucose is not taken up by most body cells and
the body begins to use fat as an energy source (ie, transfer of lipids from
adipose tissue to the liver for mobilization as an energy source). As its level is
a central metabolic control mechanism, its status is also used as a control
signal to other body systems (such as amino acid uptake by body cells). It has
several other anabolic effects throughout the body.
PRINCIPLE OF THE ASSAY
The microtiter plate provided in this kit has been pre-coated with an antibody
specific to INS. Standards or samples are then added to the appropriate
microtiter plate wells with a Horseradish Peroxidase (HRP)-conjugated
2
----------------------- Page 3-----------------------
monoclonal antibody preparation specific for INS and incubated. Then
substrate solution A and B are added to each well. Only those wells that
contain INS, HRP-conjugated antibody will exhibit a change in color. The
enzyme-substrate reaction is terminated by the addition of a sulphuric acid
solution and the color change is measured spectrophotometrically at a
wavelength of 450 nm ± 2 nm. The concentration of INS in the samples is then
determined by comparing the O.D. of the samples to the standard curve.
DETECTION RANGE
8μIU/ml-140μIU/ml. The standard curve concentrations used for the ELISA’s
were 140μIU/ml,80μIU/ml,32μIU/ml, 16μIU/ml, 8μIU/ml
SPECIFICITY
This assay recognizes recombinant and natural mouse INS. No significant
cross-reactivity or interference was observed.
SENSITIVITY
The minimum detectable dose of mouse INS is typically less than 5μIU/ml.
The sensitivity of this assay, or Lower Limit of Detection (LLD) was defined as
the lowest concentration that could be differentiated from zero.
3
----------------------- Page 4-----------------------
MATERIALS PROVIDED
Reagent Quantity
Assay plate 1
Standard(S -S ) 5
1 5
HRP-conjugate 1 x 6 ml
1 x 15 ml
Wash Buffer
(20×concentrate)
Substrate A 1 x 7 ml
Substrate B 1 x 7 ml
Stop Solution 1 x 7 ml
Standard Standard Standard Standard Standard
Standard
1 2 3 4 5
Concentration
8 16 32 80 140
(μIU/ml)
STORAGE
1. Unopened test kits should be stored at 2-8°C upon receipt and the
microtiter plate should be kept in a sealed bag to minimize exposure to
damp air. The test kit may be used throughout the expiration date of the kit.
Refer to the package label for the expiration date.
2. Opened test kits will remain stable until the expiring date shown, provided it
is stored as prescribed above.
4
----------------------- Page 5-----------------------
3. A microtiter plate reader with a bandwidth of 10 nm or less and an optical
density range of 0-3 OD or greater at 450nm wavelength is acceptable for
use in absorbance measurement.
REAGENT PREPARATION
1. Bring all reagents and plate to room temperature for at least 30 minutes
before use. Unused wells need store at 2-8°C and av oid sunlight.
2. Wash Buffer If crystals have formed in the concentrate, warm to room
temperature and mix gently until the crystals have compley dissolved.
Dilute 15 ml of Wash Buffer Concentrate into deionized or distilled water to
prepare 300 ml of Wash Buffer.
Precaution: The Stop Solution provided with this kit is an acid solution. Wear eye,
hand, face, and clothing protection when using this material.
OTHER SUPPLIES REQUIRED
Microplate reader capable of measuring absorbance at 450 nm, with the
correction wavelength set at 540 nm or 570 nm.
Pipettes and pipette tips.
Deionized or distilled water.
Squirt bottle, manifold dispenser, or automated microplate washer.
5
----------------------- Page 6-----------------------
SAMPLE COLLECTION AND STORAGE
Serum Use a serum separator tube (SST) and allow samples to clot for
30 minutes before centrifugation for 15 minutes at 1000 g. Remove serum
and assay immediay or aliquot and store samples at -20° C. Avoid
repeated freeze-thaw cycles.
Plasma Collect plasma using citrate, EDTA, or heparin as an
anticoagulant. Centrifuge for 15 minutes at 1000 g within 30 minutes of
collection. Assay immediay or aliquot and store samples at -20°C. Avoid
repeated freeze-thaw cycles.
Note: Grossly hemolyzed samples are not suitable for use in this assay.
ASSAY PROCEDURE
Bring all reagents and samples to room temperature before use. It is recommended
that all samples, standards, and controls be assayed in duplicate.
1. Set a Blank well without any solution. Add 50μl of Standard or Sample per
well. Standard need test in duplicate.
2. Reconstitute the every Standard with 0.5 ml of distilled water.
3. Add 50μl of HRP-conjugate to each well (not to Blank well). Mix well and
then incubate for 2 hour at 37°C.
6
----------------------- Page 7-----------------------
4. Complete remove the liquid. Then fill each well with Wash Buffer (about
200μl), stay for 10 seconds and Spinning. Repeat the process for a total of
three washes. Complete removal of liquid at each step is essential to good
performance. After the last wash, remove any remaining Wash Buffer by
aspirating or decanting. Invert the plate and blot it against clean paper
towels.
5. Add 50μl of Substrate A and 50μl of Substrate B to each well, mix well.
Incubate for 15 minutes at 37°C. Keeping the plate away from drafts and
other temperature fluctuations in the dark.
6. Add 50μl of Stop Solution to each well. If color change does not appear
uniform, gently tap the plate to ensure thorough mixing.
7. Determine the optical density of each well within 10 minutes, using a
microplate reader set to 450 nm.
CALCULATION OF RESULTS
Average the duplicate readings for each standard, control, and sample and
subtract the average zero standard optical density. Create a standard curve by
reducing the data using computer software capable of generating a four
parameter logistic (4-PL) curve-fit. As an alternative, construct a standard
curve by plotting the mean absorbance for each standard on the y-axis against
7
----------------------- Page 8-----------------------
the concentration on the x-axis and draw a best fit curve through the points on
the graph. The data may be linearized by plotting the log of the INS
concentrations versus the log of the O.D. and the best fit line can be
determined by regression analysis. This procedure will produce an adequate
but less precise fit of the data. If samples have been diluted, the concentration
read from the standard curve must be multiplied by the dilution factor.
LIMITATIONS OF THE PROCEDURE
The kit should not be used beyond the expiration date on the kit label.
Do not mix or substitute reagents with those from other lots or sources.
It is important that the Calibrator Diluent selected for the standard curve be
consistent with the samples being assayed.
If samples generate values higher than the highest standard, dilute the
samples with the appropriate Calibrator Diluent and repeat the assay.
Any variation in Standard Diluent, operator, pipetting technique, washing
technique, incubation time or temperature, and kit age can cause variation
in binding.
This assay is designed to eliminate interference by soluble receptors,
binding proteins, and other factors present in biological samples. Until all
factors have been tested in the Quantikine Immunoassay, the possibility of
interference cannot be excluded.
8
----------------------- Page 9-----------------------
TECHNICAL HINTS
When mixing or reconstituting protein solutions, always avoid foaming.
To avoid cross-contamination, change pipette tips between additions of
each standard level, between sample additions, and between reagent
additions. Also, use separate reservoirs for each reagent.
When using an automated plate washer, adding a 30 second soak period
following the addition of wash buffer, and/or rotating the plate 180 degrees
between wash steps may improve assay precision.
To ensure accurate results, proper adhesion of plate sealers during
incubation steps is necessary.
Substrate Solution should remain colorless until added to the plate. Keep
Substrate Solution protected from light. Substrate Solution should change
from colorless to gradations of blue.
Stop Solution should be added to the plate in the same order as the
Substrate Solution. The color developed in the wells will turn from blue to
yellow upon addition of the Stop Solution. Wells that are green in color
indicate that the Stop Solution has not mixed thoroughly with the Substrate
Solution.
9
----------------------- Page 10-----------------------
小鼠小鼠胰島素胰島素 ((Insulin))ELISA 快速檢測試劑盒快速檢測試劑盒
小小鼠鼠胰島素胰島素 (( )) 快速檢測試劑盒快速檢測試劑盒
使用說明書使用說明書
使用說明書使用說明書
本試劑盒僅供研究使用本試劑盒僅供研究使用
本試劑盒僅供研究使用本試劑盒僅供研究使用
產品編號產品編號:CSB-E05071m
產品編號產品編號
檢測范圍檢測范圍::8μIU /ml -140μIU /ml
檢測范圍檢測范圍::
特異性特異性::本試劑盒可同時檢測天然或重組的 INS,且與其他相關蛋白基本無
特異性特異性::
交叉反應。
有效期有效期:6 個月 (2-8℃避光保存)
有效期有效期
預期應用預期應用::ELISA 法定量測定小鼠血清,血漿及其它相關生物液體中INS 含
預期應用預期應用::
量。
說明說明
說明說明
1. 濃洗滌液低溫保存會有鹽析出,稀釋時可在水浴中加溫助溶。
2. 剛開啟的酶聯板孔中可能會含有少許水樣物質,此為正常現象,不會對實驗
結果造成任何影響。
實驗原理實驗原理
實驗原理實驗原理
采用酶聯免疫法夾心法檢測胰島素含量。首先用一株單抗體包被微孔板,
制備成固相抗體,然后加入待測標本及辣根過氧化物酶標記的另一株單抗,使
之形成包被抗體-胰島素-酶標記抗體的復合物。經顯色后在酶標儀測定吸光值
(OD 值),通過計算機或作圖擬合濃度-吸光度曲線,反算出待測標本中胰島素
含量。
10
----------------------- Page 11-----------------------
試劑盒組成及試劑配制試劑盒組成及試劑配制
試劑盒組成及試劑配制試劑盒組成及試劑配制
1. 酶聯板酶聯板(Assay plate ): 一塊(96 孔)。
酶聯板酶聯板
2. 標準品標準品(Standard): 5 瓶(凍干品)。
標準品標準品
Standard 1 Standard 2 Standard 3 Standard 4 Standard 5
8μIU/ml 16μIU/ml 32μIU/ml 80μIU/ml 140μIU/ml
3. 酶結合物酶結合物((HRP-conjugate):) 1×6ml/瓶。
酶結合物酶結合物(( ))
4. 顯色劑顯色劑A ((Substrate A):): 1×7ml/瓶。
顯色劑顯色劑 (( ):):
5. 顯色劑顯色劑B ((Substrate B):): 1×7ml/瓶。
顯色劑顯色劑 (( ):):
6. 濃洗滌液濃洗滌液((Wash Buffer ):) 1×15ml/瓶,使用時每瓶用蒸餾水稀釋20 倍。
濃洗滌液濃洗滌液(( ))
7. 終止液終止液((Stop Solution):) 1×7ml/瓶
終止液終止液(( ))
需要而未提供的試劑和器材需要而未提供的試劑和器材
需要而未提供的試劑和器材需要而未提供的試劑和器材
1. 標準規格酶標儀
2. 高速離心機
3. 電熱恒溫培養箱
4. 干凈的試管和Eppendof 管
5. 系列可調節移液器及吸頭,一次檢測樣品較多時,用多通道移液器
6. 蒸餾水,容量瓶等
標本的標本的采集及保存采集及保存
標本的標本的采集及保存采集及保存
1. 血清:全血標本請于室溫放置2 小時或4℃過夜后于1000 x g 離心20 分鐘,
取上清即可檢測,或將標本放于-20℃或-80℃保存,但應避免反復凍融。
2. 血漿:可用EDTA 或肝素作為抗凝劑,標本采集后30 分鐘內于2 - 8° C 1000
x g 離心15 分鐘,或將標本放于-20℃或-80℃保存,但應避免反復凍融。
3. 細胞培養物上清或其它生物標本:1000 x g 離心20 分鐘,取上清即可檢測,
或將標本放于-20℃或-80℃保存,但應避免反復凍融。
注:注:以上標本置以上標本置4℃℃保存應小于保存應小于1 周,周,-20℃℃或或-80℃℃均應密均應密封保存封保存,,-20℃℃不應超過不應超過1 個月個月,,
注注::以上標本置以上標本置 ℃℃保存應小于保存應小于 周周,, ℃℃或或 ℃℃均應密均應密封保存封保存,, ℃℃不應超過不應超過 個月個月,,
-80℃℃不應超過不應超過2 個月個月;標本溶血會影響zui后檢測結果;標本溶血會影響zui后檢測結果,因此溶血標本不宜進行此項檢測,因此溶血標本不宜進行此項檢測。。
℃℃不應超過不應超過 個月個月;;標本溶血會影響zui后檢測結果標本溶血會影響zui后檢測結果,,因此溶血標本不宜進行此項檢測因此溶血標本不宜進行此項檢測。。
11
----------------------- Page 12-----------------------
操作步驟操作步驟
操作步驟操作步驟
1. 將各種試劑至室溫 〔18-25℃〕平衡半小時,取濃縮洗滌液,根據當批檢測
數量,用蒸餾水上1:20 稀釋,混勻后備用。
2. 標準品S1- S 5,*次使用前先用0.5ml 蒸餾水溶解,放置混勻后使用。
3. 將酶標板取出,設一個空白對照孔、不加任何液體;每個標準點依次各設兩
孔,每孔加入相應標準品50ul;其余每個檢測孔直接加待測標本50ul。
4. 每孔加入酶結合物50ul (空白對照孔除外),充分混勻,貼上不干膠封片,
置37℃溫育2 小時。
5. 手工洗板,棄去孔內液體。洗滌液注滿各孔,靜置10 秒甩干,重復三次后
拍干;洗板機洗板,選擇洗滌三次程序,洗板后拍干。
6. 每孔加顯色劑A 液50μl,顯色劑B 液50μl,振蕩混勻后,37℃避光顯色15
分鐘,每孔加終止液50μl。
7. 用酶標儀讀數,取波長450nm,先用空白孔調零點,然后測定各孔OD 值。
數據處理數據處理
數據處理數據處理
1. 手工作圖:用雙對數坐標紙,以標準品濃度為橫軸,以對應的0D 值為縱軸,
畫出平滑曲線或直線,在曲線上按照待測血清OD 值找到對應的濃度值。
2. 計算機:使用線性擬合功能,應將標準品S1-S5 的濃度取對數(Log(濃度))
作為X,將對應的OD 值減去空白對照孔OD 值后取對數(Log(OD 值-NSB))
作為Y,進行線性擬合。再從擬合線上計算出待測血清濃度。
注意事項注意事項
注意事項注意事項
1. 從冷藏環境中取出的試劑盒內全部瓶裝試劑及所需預包被板條應置室溫
(18-25℃)平衡30 分鐘后方可使用,余者應及時封好口,放回2-8℃中避光保
存,以備后用。
2. 使用前試劑應搖勻。
3. 結果判斷須在反應終止后10 分鐘內完成。
4. 不同批號的試劑不可混用。
5. 加樣時應注意避免所用各試劑及樣品之間的交又污染。
6. 操作時,試劑盒內每種試劑各使用一個吸頭,每一種標準品使用一個吸頭,
每一個樣品各使用一個吸頭,吸頭一次性使用。
12
----------------------- Page 13-----------------------
Notes
13
----------------------- Page 14-----------------------
14
----------------------- Page 15-----------------------
15
----------------------- Page 16-----------------------
16
----------------------- Page 17-----------------------
17
環保在線 設計制作,未經允許翻錄必究 .? ? ?
請輸入賬號
請輸入密碼
請輸驗證碼
請輸入你感興趣的產品
請簡單描述您的需求
請選擇省份