  Mouse Leptin  ((LEP))ELISA kit
                                     ((  ))
 
                      Catalog No. CSB-E04650m
 
                                    (96T)
 
    This immunoassay kit allows for the in vitro quantitative determination of mouse
 
    LEP   concentrations  in cell  culture  supernates,  serum,  plasma  and  other
 
    biological fluids.
 
    Expiration date   six months from the date of manufacture
 
    FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
 
                       CUSABIO BIOTECH CO., Ltd.
 
              /     /
 
           : cusabio@        cusabio@
 
                                       1


----------------------- Page 2-----------------------
INTRODUCTION
 
Leptin   (from   the   Greek   leptos,   meaning   thin)   is   a   protein   hormone   with
 
important   effects   in   regulating   body   weight,   metabolism   and   reproductive
 
function. The protein is approximay ~16 kDa in mass and encoded by the
 
obese (ob) gene. Leptin is expressed predominantly by adipocytes, which
 
fits with the idea that body weight is sensed as the total mass of fat in the
 
body. Smaller amounts of leptin are also secreted by cells in the epithelium
 
of the stomach and in the placenta. Leptin receptors are highly expressed in
 
areas of the hypothalamus known to be important in regulating body weight,
 
as well as in T lymphocytes and vascular endothelial cells.
 
PRINCIPLE OF THE ASSAY
 
The    microtiter   plate   provided    in  this  kit  has  been   pre-coated     with   an
 
antibody   specific   to   LEP.   Standards   or   samples   are  then   added   to   the
 
appropriate      microtiter   plate   wells   with   a   biotin-conjugated      polyclonal
 
antibody preparation
 
specific for LEP. and Avidin conjugated to Horseradish Peroxidase (HRP) is
 
added     to  each    microplate    well  and   incubated.    Then    a   TMB    (3,3'5,  5'
 
tetramethyl-benzidine) substrate solution is added to each well. Only those
 
wells that contain LEP., biotin-conjugated antibody and enzyme-conjugated
 
Avidin   will   exhibit   a   change   in   color.   The   enzyme-substrate   reaction   is
 
terminated by the addition of a sulphuric acid solution and the color change
 
is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The
 
concentration of LEP. in the samples is then determined by comparing the
 
O.D. of the samples to the standard curve.
 
                                             2

----------------------- Page 3-----------------------
DETECTION RANGE
 
0.16   ng/ml-10   ng/ml.   The  standard   curve   concentrations   used   for  the
 
ELISA’s   were   10   ng/ml,   5   ng/ml,   2.5   ng/ml,   1.25   ng/ml,   0.62   ng/ml,   0.32
 
ng/ml, 0.16 ng/ml
 
SPECIFICITY
 
This assay recognizes recombinant and natural mouse LEP No significant
 
cross-reactivity or interference was observed.
 
SENSITIVITY
 
The   minimum   detectable   dose   of   mouse   LEP   is   typically   less   than   0.04
 
ng/ml.
 
The sensitivity of this assay, or Lower Limit of Detection (LLD) was defined
 
as the lowest protein concentration that could be differentiated from zero.
 
MATERIALS PROVIDED
 
              Reagent                                    Quantity
              Assay plate                                     1
              Standard                                        2
              Sample Diluent                             1 x 20 ml
              Biotin-antibody Diluent                    1 x 10 ml
              HRP-avidin Diluent                         1 x 10 ml
              Biotin-antibody                            1 x 120μl
              HRP-avidin                                 1 x 120μl
                                                         1 x 20 ml
              Wash Buffer
                                                      (25×concentrate)
              TMB Substrate                              1 x 10 ml
              Stop Solution                              1 x 10 ml
 
                                        3

----------------------- Page 4-----------------------
STORAGE
 
1.   Unopened   test   kits   should   be   stored   at   2-8°C   upon   receipt   and   the
 
    microtiter    plate  should    be   kept  in  a  sealed    bag   with  desiccants     to
 
    minimize exposure to damp air. The test kit may be used throughout the
 
    expiration date of the kit. Refer to the package label for the expiration
 
    date.
 
2.   Opened      test  kits  will  remain   stable   until  the  expiring   date   shown,
 
    provided it is stored as prescribed above.
 
3.   A   microtiter   plate   reader   with   a   bandwidth   of   10   nm   or   less   and   an
 
    optical   density   range   of   0-3   OD   or   greater   at   450nm   wavelength   is
 
    acceptable for use in absorbance measurement.
 
REAGENT PREPARATION
 
Bring all reagents to room temperature before use.
 
1.   Wash Buffer         If crystals have formed in the concentrate, warm up to
 
     room   temperature   and   mix   gently   until   the   crystals  have   compley
 
     dissolved. Dilute 20 ml of Wash Buffer Concentrate  into deionized or
 
     distilled water to prepare 500 ml of Wash Buffer.
 
2.   Standard       Reconstitute the Standard with 1.0 ml of Sample Diluent.
 
     This   reconstitution   produces   a   stock   solution   of   10   ng/ml.   Allow   the
 
     standard to sit for a minimum of 15 minutes with gentle agitation prior to
 
     making   serial    dilutions.   The   undiluted   standard     serves   as  the   high
 
     standard (10 ng/ml). The Sample Diluent serves as the zero standard
 
     (0 ng/ml).
 
                                             4

----------------------- Page 5-----------------------
3.   Biotin-antibody       Dilute to the working concentration specified on the
 
     vial label using Biotin-antibody Diluent(1:100), respectively.
 
4.   HRP-avidin       Dilute to the working concentration specified on the vial
 
     label using HRP-avidin Diluent(1:100), respectively.
 
Precaution: The Stop Solution provided with this kit is an acid solution. Wear
 
            eye, hand, face, and clothing protection when using this material.
 
OTHER SUPPLIES REQUIRED
 
     Microplate   reader   capable   of   measuring   absorbance  at   450   nm,   with
 
     the correction wavelength set at 540 nm or 570 nm.
 
     Pipettes and pipette tips.
 
     Deionized or distilled water.
 
     Squirt bottle, manifold dispenser, or automated microplate washer.
 
SAMPLE COLLECTION AND STORAGE
 
     Cell Culture Supernates         Remove particulates by centrifugation and
 
     assay immediay or aliquot and store samples at -20° C. Avoid
 
     repeated freeze-thaw cycles.
 
     Serum     Use a serum separator tube (SST) and allow samples to clot
 
     for 30 minutes before centrifugation for 15 minutes at 1000 g. Remove
 
     serum and assay immediay or aliquot and store samples at -20° C.
 
    Avoid repeated freeze-thaw cycles.
 
     Plasma     Collect plasma using citrate, EDTA, or heparin as an
 
     anticoagulant. Centrifuge for 15 minutes at 1000 x g within 30 minutes
 
     of collection. Assay immediay or aliquot and store samples at -20° C.
 
    Avoid repeated freeze-thaw cycles.
 
Note: Grossly hemolyzed samples are not suitable for use in this assay.
 
                                          5

----------------------- Page 6-----------------------
ASSAY PROCEDURE
 
Bring all reagents and samples to room temperature before use. It is
 
recommended that all samples, standards, and controls be assayed in duplicate.
 
1.    Add   100μl   of   Standard,   Blank,   or   Sample   per   well.   Cover   with   the
 
     adhesive strip. Incubate for 2 hours at 37° C.
 
2.    Remove the liquid of each well, don’t wash.
 
3.    Add 100μl of Biotin-antibody working solution to each well. Incubate
 
     for   1  hour   at  37°C.   Biotin-antibody   working         solution   may    appear
 
     cloudy.   Warm   up   to   room   temperature   and   mix   gently  until   solution
 
     appears uniform.
 
4.    Aspirate each well and wash, repeating the process three times for a
 
     total of three washes. Wash by filling each well with Wash Buffer (200μl)
 
     using     a  squirt   bottle,  multi-channel     pipette,   manifold     dispenser     or
 
     autowasher.   Complete   removal   of   liquid   at   each   step   is   essential   to
 
     good   performance.   After   the   last   wash,   remove   any   remaining Wash
 
     Buffer   by   aspirating   or   decanting.   Invert   the   plate  and   blot   it   against
 
     clean paper towels.
 
5.    Add   100μl   of  HRP-avidin   working   solution   to   each   well.   Cover   the
 
     microtiter plate with a new adhesive strip. Incubate for 1 hour at 37° C.
 
6.    Repeat the aspiration and wash three times as step 4.
 
7.    Add 90μl of TMB Substrate to each well. Incubate for 30 minutes at
 
     37°C.    Keeping     the   plate   away    from   drafts   and   other   temperature
 
     fluctuations in the dark.
 
8.    Add   50μl   of  Stop   Solution   to   each   well.   If   color   change   does   not
 
     appear uniform, gently tap the plate to ensure thorough mixing.
 
9.    Determine the optical density of each well within 30 minutes, using a
 
     microplate reader set to 450 nm.
 
                                              6

----------------------- Page 7-----------------------
CALCULATION OF RESULTS
 
Average the duplicate readings for each standard, control, and sample and
 
subtract the average zero standard optical density. Create a standard curve
 
by reducing the data using computer software capable of generating a four
 
parameter   logistic   (4-PL)curve-fit.   As   an   alternative,   construct   a   standard
 
curve   by   plotting   the   mean   absorbance   for   each   standard   on   the   y-axis
 
against the concentration on the x-axis and draw a best fit curve through the
 
points on the graph. The data may be linearized by plotting the log of the
 
LEP. concentrations versus the log of the O.D. and the best fit line can be
 
determined       by   regression     analysis.   This    procedure     will  produce     an
 
adequate but less precise fit of the data. If samples have been diluted, the
 
concentration      read   from   the  standard    curve    must   be  multiplied    by  the
 
dilution factor.
 
LIMITATIONS OF THE PROCEDURE
 
      The kit should not be used beyond the expiration date on the kit label.
 
      Do not mix or substitute reagents with those from other lots or sources.
 
      It is important that the Calibrator Diluent selected for the standard curve
 
     be consistent with the samples being assayed.
 
      If samples generate values higher than the highest standard, dilute the
 
     samples with the appropriate Calibrator Diluent and repeat the assay.
 
      Any    variation   in  Standard     Diluent,    operator,    pipetting   technique,
 
     washing   technique,   incubation   time   or   temperature,  and   kit   age   can
 
     cause variation in binding.
 
      This assay is designed to eliminate interference by soluble receptors,
 
     binding proteins, and other factors present in biological samples. Until
 
     all  factors   have    been    tested   in  the  Quantikine     Immunoassay,       the
 
     possibility of interference cannot be excluded.
 
                                             7

----------------------- Page 8-----------------------
TECHNICAL HINTS
 
      When mixing or reconstituting protein solutions, always avoid foaming.
 
      To avoid cross-contamination, change pipette tips between additions of
 
     each standard level, between sample additions, and between reagent
 
     additions. Also, use separate reservoirs for each reagent.
 
      When   using   an   automated   plate   washer,   adding   a   30  second   soak
 
     period   following   the   addition   of   wash   buffer,   and/or   rotating   the   plate
 
     180 degrees between wash steps may improve assay precision.
 
      To   ensure   accurate   results,   proper   adhesion   of   plate   sealers   during
 
     incubation steps is necessary.
 
      Substrate   Solution   should   remain   colorless   until   added   to   the   plate.
 
     Keep Substrate Solution protected from light. Substrate Solution should
 
     change from colorless to gradations of blue.
 
      Stop Solution should be added to the plate in the  same order as the
 
     Substrate Solution. The color developed in the wells will turn from blue
 
     to   yellow   upon   addition   of   the   Stop   Solution. Wells  that   are   green   in
 
     color indicate that the Stop Solution has not mixed thoroughly with the
 
     Substrate Solution.
 
                                             8
 

 

環保在線 設計制作,未經允許翻錄必究 .? ? ? Copyright(C)?2021 http://www.aboay.com,All rights reserved.

以上信息由企業自行提供,信息內容的真實性、準確性和合法性由相關企業負責,環保在線對此不承擔任何保證責任。 溫馨提示:為規避購買風險,建議您在購買產品前務必確認供應商資質及產品質量。

會員登錄

×

請輸入賬號

請輸入密碼

=

請輸驗證碼

收藏該商鋪

登錄 后再收藏

提示

您的留言已提交成功!我們將在第一時間回復您~
主站蜘蛛池模板: 深水埗区| 温州市| 英山县| 夹江县| 临颍县| 育儿| 天全县| 蓬莱市| 达孜县| 襄樊市| 马尔康县| 绵竹市| 黔东| 忻城县| 莒南县| 石河子市| 黑水县| 惠安县| 益阳市| 新营市| 漳州市| 南城县| 潼南县| 塔城市| 滁州市| 永泰县| 泗水县| 景德镇市| 南丰县| 南华县| 通城县| 策勒县| 龙州县| 西贡区| 保康县| 陆丰市| 安岳县| 秭归县| 海阳市| 霍城县| 渑池县|